CRSY102 Operations Manual

Rev G 1/31/2013

Page 26

SECTION 6

– BRIEF TROUBLESHOOTING

Reservoir viewing window does not illuminate (blue)
 Is recirculator plugged in?
 Is the power switch turned on?
 Has the recirculator circuit breaker tripped? If so, reset internal circuit breaker.
 Is there water in the reservoir? Add water light will be on.







Water is leaking from unit
 Is the recirculator on?
 Has a power failure recently occurred?
 Are the filter housings tight?
 Is the o-ring in the filter housing properly seated?
 Is the internal reservoir drain valve open?
 Is the recirculator water outlet connected properly to the chamber?


Water is not draining properly from the chamber(s) to the recirculator
 Is the recirculator physically lower than the chamber(s)?
 Is the chamber drain line (recirculator inlet) kinked? Plugged? Have a rise?
 Is the recirculator off?


It is part of the normal operation for
 The blue reservoir light to flicker
 Water to flow into the reservoir producing a splashing and stirring motion
 The red Replace Filters indicator light to flash at startup



The recirculator must be powered on to prevent
returning condensate (water from chamber) from
overflowing inside the recirculator.