CRSY102 Operations Manual

Rev G 1/31/2013

Page 16

Starting Recirculator

1. Turn recirculator power switch on.

Reservoir viewing window will illuminate (blue)

Replace filter indicator light (red) will blink once

Internal air purging process will begin

Recirculator outlet will become pressurized

Internal water purification cycle will initiate (as necessary)

2. Purge air from chamber(s) water supply line (recirculator outlet) by setting the

chamber(s) to high temperature and high humidity set points.





3. Wait a few minutes as water fills the tubing in the chamber
4. Verify the humidity level has increased inside the chamber
5. Set temperature & humidity set points on chamber(s) to desired settings.








Allow the chamber(s) to dry out before turning
any internal lights on.

For chamber(s) with internal lights, turn the
lights off.