– RECIRCULATOR STARTUP



















Filling Reservoir

1. Unscrew reservoir cap.
2. Fill reservoir with water (distilled, low-grade deionized or tap). Reservoir holds

7 gallons (26.5 liters).

3. Screw reservoir cap back in place.

Be sure recirculator outlet is connected before
unit is turned on. As soon as the recirculator is
turned on, water will begin flowing from the outlet.

When adding water to the reservoir, do not spill
onto recirculator.

Ultra pure DI water is chemically aggressive. Do
not put DI water with a resistivity above 1

MΩ-cm

in the recirculator.