Temperature – Despatch RBC1-50 with Protocol Plus User Manual

Page 10

Advertising
background image

6

Temperature

RBC1-50

Air Atmosphere

RBC1-50

Nitrogen Atmosphere

Time to Temperature

25

°

C - 80

°

C

(approximate minutes with

25

°

C - 125

°

C

no live load)

(1)

25

°

C - 150

°

C

25

°

C - 260

°

C

4

10
14
34

4

10
14
30

Recovery Time to 125

°

C with door open 1 Min.

>1 Min.

>1 Min.

Temperature Uniformity at

(2)

125

°

C*

260

°

C*

<±1°C

±1°C

<±1°C

±1°C

Operating Temperature Range
(Based on 25

°

C ambient, and dampers fully open)

50° - 260°C

15° - 260°C

Control Stability

(3)

±0.5°C per 5°C

∆

Repeatability

±0.5°C

Cooldown Time 125° - 60

°

C

40 Min.

11 Min.

Dissipation/Mean Temperature Rise

(4)

80°C

125°C
150°C

800 Watts/4°C

1,300 Watts/6°C
1,400 Watts/6°C


(1)

Time to temperature will be slightly longer for lower voltages due to heater kW derating (empty oven).

(2)

Figures are based on a nine-point survey in an empty oven. Uniformity can vary slightly depending

on unit and operating conditions.

(3)

∆

In control stability figures represent the change in ambient temperature.

(4)

Dissipation is reduced by 17% on 50-Hertz operation. Temperature gradient/mean temperature rise is

with an evenly distributed live load.

Note: Testing performed on 240V-1PH-60HZ.

Advertising