Despatch RBC1-50 with Protocol Plus User Manual

Page 38

Advertising
background image

34

Difficulty

Probable Cause

Suggested Remedy

Improper airflow

Fan motor failure

Fan wheel seated too low on
fan shaft

Unbalanced fan wheel

Replace fan motor.

Adjust fan wheel for 3/16" clearance between
wheel and inlet ring.

Replace fan wheel.

Excessive
vibration

Dirty fan wheel

Unbalanced fan wheel

Clean fan.

Replace fan wheel.

Oven will not
control at
setpoint

High limit set too low

Control malfunction

SSR malfunction

Air friction of recirculation fan


Damper motor (exhaust) not
responding

Set the high limit higher

Replace control.

Replace SSR and/or check control output voltage.

Open fresh air vent. Unit will not control below
minimum operating temperature with vent closed.

Check tuning parameters.

Damper motor
(exhaust) not
responding

Linkage binding

Motor not working

Improper tuning parameters

Observe and correct linkage.

Check supply and signal voltage.

Check tuning parameters.

Liquid nitrogen
(LN

2

) not cooling

Solenoid valve not responding


Improper tuning parameters

Tank empty or valve is to the
OFF position

Pressure regulator set too low

Check supply voltage and control output. Replace
valve.

Check tuning parameters.

Check liquid nitrogen (LN

2

) supply and valve

position.

Adjust regulator to 35PSIG maximum.

Heater does not
shut down until
temp. reaches
the High limit
setting

Control malfunction

SSR malfunction

Replace control

Replace SSR

Nitrogen (N

2

) not

flowing correctly

Tank empty or valve is to the
OFF position

Pressure regulator set too low


Flowmeter incorrectly set

Check nitrogen supply and valve position.


Adjust regulator within the acceptable range of the
flowmeter.

Adjust flowmeter to the desired flow rate.