Despatch RBC1-50 with Protocol Plus User Manual
Page 38

34
Difficulty
Probable Cause
Suggested Remedy
Improper airflow
Fan motor failure
Fan wheel seated too low on
fan shaft
Unbalanced fan wheel
Replace fan motor.
Adjust fan wheel for 3/16" clearance between
wheel and inlet ring.
Replace fan wheel.
Excessive
vibration
Dirty fan wheel
Unbalanced fan wheel
Clean fan.
Replace fan wheel.
Oven will not
control at
setpoint
High limit set too low
Control malfunction
SSR malfunction
Air friction of recirculation fan
Damper motor (exhaust) not
responding
Set the high limit higher
Replace control.
Replace SSR and/or check control output voltage.
Open fresh air vent. Unit will not control below
minimum operating temperature with vent closed.
Check tuning parameters.
Damper motor
(exhaust) not
responding
Linkage binding
Motor not working
Improper tuning parameters
Observe and correct linkage.
Check supply and signal voltage.
Check tuning parameters.
Liquid nitrogen
(LN
2
) not cooling
Solenoid valve not responding
Improper tuning parameters
Tank empty or valve is to the
OFF position
Pressure regulator set too low
Check supply voltage and control output. Replace
valve.
Check tuning parameters.
Check liquid nitrogen (LN
2
) supply and valve
position.
Adjust regulator to 35PSIG maximum.
Heater does not
shut down until
temp. reaches
the High limit
setting
Control malfunction
SSR malfunction
Replace control
Replace SSR
Nitrogen (N
2
) not
flowing correctly
Tank empty or valve is to the
OFF position
Pressure regulator set too low
Flowmeter incorrectly set
Check nitrogen supply and valve position.
Adjust regulator within the acceptable range of the
flowmeter.
Adjust flowmeter to the desired flow rate.