The transformer is a not switch able transformer, so the main voltage must match to
the one marked on the front cover. If the main power is different then this can
damage the transformer.
In the ODIU there can be found four fuses, two for the transformer and the other two
for the head.

Fuses

On the primarily side, the fuse ratings are: T1,6A (250V)

- in the case of glassheater the fuses are T1,6A (250V)

On the secondary side, the fuse ratings are: T10A (250V)

On the glassheater-panel the fuses are T6,3A (250V)


Installing the ODIU box


The physical fixing of the ODIU box can be seen below.

Necessary toolkit:

• Philips

screwdriver

• 5mm

wrench

• Drilling machine with an Ø8mm bit (the minimum effective length is 50mm)

First you have to fix the ODIU box to the metal plate:

• Pick the four M6x70 flat sphere headed screws, and take them to the holes

on the corners of the box

• On the back of the box, where the screws came out, you have to put up the

four 6x40 spacers (these are the longer ones) to the legs of the screws

• Pull on the metal plate to the legs of the screws

• Fix the metal plate with four M5 nuts using the four washers, and four spring

washers like in the figure













Then you can fix the whole construction to a wall: