8 Errors and Troubleshooting

8.1 General errors

Error / message

Cause

Remedy

Unit cannot be switched on

Main plug not connected to the main pow-
er supply

 Connect the main plug

with the power supply

Fuses defective or blown fuse

 Replace fuses

(see chapter 9.2)

No heating function

Master switch is deactivated

 Activate main switch

Fuses defective or blown fuse

 Replace fuses

(see chapter 9.2)

Master switch is defective

 Contact service

The connecting cable of the heating bath is
not connected.

 Connect the connecting

cable of the heating bath

The heater of the heating bath is defective

 Contact service

The maximum temperature limiter re-
sponded

 If the medium was in the

heating bath: Let the
heating bath cool and re-
set the maximum tem-
perature limiter

 If the medium was not in

the heating bath: Contact
service

Drive unit is not rotating

Master switch is defective

 Contact service

Speed controller is on the left stop
(Hei-VAP value)

 Turn the speed controller

to the right

Drive unit is defective

 Contact service

Fuses defective or blown fuse

 Replace fuses

(see chapter 9.2)

Motor lift is not functioning

Master switch is deactivated

 Activate main switch

Master switch is defective

 Contact service

Fuses defective or blown fuse

 Replace fuses

(see chapter 9.2)

Lift is at end stop

 Activate other arrow key

Mechanical system / motor is defective

 Contact service

Height stop is not set accurately

 Perform height adjust-

ment