Hei-vap value, 3 hei-vap value – Heidolph Hei-VAP Advantage User Manual

Page 62

Advertising
background image

Errors and Troubleshooting

54

Hei-VAP Value, Hei-VAP Advantage

EN

EN

EN

8.3 Hei-VAP Value

Error / message

(flashing code)

Cause

Remedy

**

No data received at all for 3 se-
conds

 Cable connection between the control

panel and the base unit was discon-
nected. An error occurred in the data
bus.

 Restore the cable con-

nection between the con-
trol panel and the base
unit. Switch the power
switch off and on again.

 Contact service

***

Heater is defective, i.e. the over-
heating protection was triggered.

 If the medium was in the

heating bath: Let the
heating bath cool and re-
set the maximum tem-
perature limiter

 If the medium was not in

the heating bath: Contact
service

HB safety temperature (+5 K) is
exceeded

 Contact service

Heating bath is completely evapo-
rated.

 Deactivate the unit
 Check if the maximum

temperature limiter has
responded. If yes, reset
(see chapter 9.1.1)

 Fill in heat transfer fluid

Heating bath sensor variance
> 10 K

 Contact service

Sensor monitoring via hardware
HB sensor broken or short circuit-
ed

 Contact service

The temperature sensor for moni-
toring the heating bath control is
broken or the temperature of the
medium in the heating bath is
> 217 °C.

 Cool down the heating

bath medium

 Contact service

The temperature sensor for the
heating bath control is short cir-
cuited or the temperature is below
0 °C

 Contact service

****

Potentiometer temp. is defective
(value)

 Contact service

Potentiometer speed is defective
(value)

 Contact service

*****

Motor is not rotating:

Motor overload shutdown

 Deactivate the unit
 Stop drive unit for cooling

down

 Contact service

Table 8-3: Troubleshooting table - Hei-VAP Value