Errors and Troubleshooting


Hei-VAP Value, Hei-VAP Advantage

53

EN

Error / message

Cause

Remedy

No evacuation

Power switch Rotavac vario control or
Rotavac vario tec is not activated.

Vacuum valve is defective.

 Activate the power switch

Fuses defective or blown fuse

 Replace fuses

(see chapter 9.2)

Insufficient vacuum

System leaks

 Check seals, tubing and

connections

 Check stopcocks, grease

if necessary

Vacuum pump is defective

 Observe manufacturer's

instructions of the vacu-
um pump

Unit suddenly switches off
(Hei-VAP Precision and
Hei-VAP Advantage)

Timer is programmed

 Check and switch off the

timer, if necessary

(see chapter 7.4.1)

Fuses defective or blown fuse

 Replace fuses

(see chapter 9.2)

Table 8-1: Troubleshooting table

8.2 Additional conditions

Additional conditions

Reaction

Vapor sensor is defective

The actual value curve is displayed “---“ instead of the vapor tem-
perature.

The vapor sensor is not con-
nected

The vapor temperature is not displayed in the actual value display

Heater can only be activated at
>0 °C (observe acetone, dry
ice)

Table 8-2: Additional conditions