Hei-vap advantage, 4 hei-vap advantage – Heidolph Hei-VAP Advantage User Manual

Page 63

Advertising
background image

Errors and Troubleshooting


Hei-VAP Value, Hei-VAP Advantage

55

EN

8.4 Hei-VAP Advantage

Error / message

(error label)

Cause

Remedy

1

The transportation protection device is still
activated.

(only motor lift version)

 Please raise the lift and

briefly switch the unit off
and then on again.

No data received at all for 3 se-
conds

 Cable connection between the control

panel and the base unit was discon-
nected. An error occurred in the data
bus.

 Restore the cable con-

nection between the con-
trol panel and the base
unit. Switch the power
switch off and on again.

 Contact service

3

Heater is defective, i.e. the over-
heating protection was triggered.

 If the medium was in the

heating bath: Let the
heating bath cool and re-
set the maximum tem-
perature limiter

 If the medium was not in

the heating bath: Contact
service

Heating bath safety temperature
(+5 K) is exceeded

 Contact service

Heating bath is completely evapo-
rated.

 Deactivate the unit
 Check if the maximum

temperature limiter has
responded. If yes, reset
(see chapter 9.1.1)

 Fill in heat transfer fluid

Heating bath sensor variance
> 10 K

 Contact service

Sensor monitoring via hardware
HB sensor broken or short circuit-
ed

 Contact service

The temperature sensor for moni-
toring the heating bath control is
broken or the temperature of the
medium in the heating bath is
> 217 °C.

 Cool down the heating

bath medium

 Contact service

The temperature sensor for the
heating bath control is short cir-
cuited or the temperature is below
0 °C

 Contact service

5

Motor is not rotating:

Motor overload shutdown

 Deactivate the unit
 Stop drive unit for cooling

down

 Contact service

Table8-4: Troubleshooting table - Hei-VAP Advantage


Advertising