NOTE: Read through this section before filling the hydrogen fuel cylinder.

Store the hydrogen supply tank in a well ventilated area, well away from heat or possible ignition sources.

WARNING

Hydrogen gas is a fire and explosion hazard when exposed to heat or flame.
The lower explosive limit is 4%

Directions for Attaching the Hydrogen Filling Station to the Hydrogen Supply Tank

1. Connect the hydrogen fuel cylinder filling station to the tank of hydrogen. Hydrogen is supplied with a
CGA 350 cylinder valve outlet. The filling station is supplied with the matching fitting for ease of
connection. The threads of the adaptor will tighten counterclockwise.

NOTE: Do not force the connection. Do not use Teflon tape with CGA fittings. In general, these fittings
are designed for metal to metal sealing.

Do not use adapters to connect one CGA fitting to another type of CGA fitting. If the refill adapter does
not match the outlet on your hydrogen tank, contact INFICON.

2. Tighten the hydrogen cylinder filling station onto the tank with a wrench. Do not over tighten.

WARNING

DO NOT Use an open flame to test for leaks!

WARNING

DO NOT fill the hydrogen cylinder in a hazardous location.