Backlight, User mode – INFICON MicroFID II Portable Flame Ionization Detector User Manual

Page 53

Advertising
background image

074-579-P1A

53

To turn on the pump:

1.

Press the MENU soft key, “MicroFID II Setup”, press SELECT, “Pump”, press
SELECT, “On”, then press SELECT.

To turn off the pump:

2.

Press the MENU soft key, “MicroFID II Setup”, press SELECT, “Pump”, press
SELECT, “Off”, then press SELECT.

Backlight

The Backlight function is used to switch the backlighting on and off when there is insufficient
light to read the display.

To switch the Backlighting on and off:

1. Press the MENU soft key and select

“MicroFID II Setup”.

2. Select BACKLIGHT and then press the SELECT soft key. Press the DOWN ARROW

soft key to either turn the backlight on or off.

3. Press SELECT soft key to return to the main display.

To extend the operating life of the battery pack, turn the backlighting off when it is not required.

User Mode

User mode selects one of the three logging modes. The four logging modes are Logging Off,
Location, and Interval mode.

Logging Off mode displays instantaneous readings only. The display is updated continuously
and readings are not logged.

Location mode displays instantaneous readings. The user has the option to manually store the
instantaneous reading as either a data point with a tag and a sample reading.

Interval mode displays the instantaneous reading. In interval mode data is logged at a user
selected interval value between 1 second and 999 seconds.

To access the User Mode Section:

1. Press the

MENU soft key and select “MicroFID II Setup”.

2. Select User Mode and then press the SELECT soft key. Press the UP or DOWN

ARROW soft key to choose the applicable User Mode, that is, Logging Off, Location,
Interval, or Method 21.

3. Press SELECT soft key to return to the main display.

Advertising