8. This screen will automatically appear: Press next once you attach the first span gas.
9. The MicroFID II will automatically enter the span value. If there is a problem with the span
gas process, the MicroFID II will inform you, e.g. concentration for span gas too low.
10. Once the first span gas calibration is complete, the following screen will appear:
To go back to the first calibration step, press the back key, otherwise press next.
11. This screen will appear. Keep this second value by pressing next, or change it by pressing
new.
Calibrating Span Gas1
Perform Span 2
Calibrate next
Cancel
Back
Next
Attach
span gas 1
Cancel
Next
“next” to continue
OR ‘new’ to change
2,000 ppm
Cancel
New
Next