Interval mode - data logging – INFICON MicroFID II Portable Flame Ionization Detector User Manual

Page 49

Advertising
background image

074-579-P1A

49

enter password

The MicroFID II has the ability to change log modes without erasing the log. In other words, if
the user switches User modes from Interval mode to Location mode the previous Interval log
entries are NOT erased as in older style instruments.

The MicroFID II

also operates in a ‘circular-logging’ mode. This means that the log will never fill

and stop logging but will loop back to the first entry and start logging at entry number 1.

Interval Mode - Data Logging

Interval mode logs readings at user selected intervals of 1 second to 999 seconds. The
microprocessor accumulates all readings in an averaging interval that you select, and
determines the average reading. It stores this number along with the highest priority instrument
status and the current time and date.

Figure 28. Interval time adjustment

Location Operation

– Data Logging

In the Location

mode of operation, data is logged only when the ‘accept’ soft key on the probe

trigger is pressed. The date, time, average reading at the time the trigger is pressed, along with
the current tag string that is currently on the display are also stored.

Clearing or Downloading Data

There are two options for data manipulation under the Data Log Options selection.

The first option

is the “Clear Data Log” option which deletes all readings in the MicroFID II

memory.

NOTE: Deleted information cannot be recovered. You should download the contents of the
datalogger before deleting any information.

Enter new INTERVAL (secs)

015

done