INFICON MicroFID II Portable Flame Ionization Detector User Manual

Page 70

Advertising
background image

074-579-P1A

70

12. Attach span gas as instructed.

13. The MicroFID II will automatically enter span gas 2.

14. Once the last calibration step is performed, you will see this display:

.

15. Once the Done key is pushed, the MicroFID II will automatically return to measurement
mode (see below).

Calibration Completed

Press ‘done’ key to

exit calibration mode

Done

Attach

Span Gas 2

Cancel

Next

Calibrating Span Gas 2