INFICON MicroFID II Portable Flame Ionization Detector User Manual
Page 70
Advertising
074-579-P1A
70
12. Attach span gas as instructed.
13. The MicroFID II will automatically enter span gas 2.
14. Once the last calibration step is performed, you will see this display:
.
15. Once the Done key is pushed, the MicroFID II will automatically return to measurement
mode (see below).
Calibration Completed
Press ‘done’ key to
exit calibration mode
Done
Attach
Span Gas 2
Cancel
Next
Calibrating Span Gas 2