3 crystal sensor emulator – INFICON Front Load Single Sensor User Manual

Page 86

Advertising
background image

5 - 10

PN

07

4-

15

6N

Front Load Single and Dual Sensors Operating Manual

5.1.3 Crystal Sensor Emulator

A very useful tool for rapidly evaluating the cause of a persistent Crystal Fail
message is the optional Crystal Sensor Emulator (PN 760-601-G2) used in
conjunction with a thin film deposition controller or monitor (see

Figure 5-7

).

The Crystal Sensor Emulator contains a known "good" 5.5 MHz crystal that may be
attached at various points in the measurement system to isolate the cause of a
Crystal Fail. The three connectors on the Crystal Sensor Emulator allow
connection to a BNC cable, or an in-vacuum cable, or the sensor head cavity
(removal of the shutter may be required for a Front Load Dual sensor to allow
adequate clearance). (See

Figure 5-8

.)

To use the Crystal Sensor Emulator to isolate the cause of a Crystal Fail message,
start at

section 5.1.3.1 on page 5-11

.

NOTE: Crystal Sensor Emulator PN 760-601-G1 (obsolete) is not compatible with

IC6, IC/5, and IC/4. However, PN 760-601-G2 is fully compatible with all
INFICON thin film deposition controllers and monitors.

The value of Crystal Life (Health) displayed by the controller or monitor
depends on the model. See

section 5.1.3.4, Crystal Life Readings, on page

5-13

.

It is normal for Unable To Auto Z to be displayed if the Crystal Sensor Emulator
is used with a deposition controller configured to use the Auto Z feature.

CAUTION

Crystal Sensor Emulator is not intended for use in
vacuum. Do not leave it installed in the vacuum system
during processing.

Figure 5-7 Crystal Sensor Emulator

Female

BNC

Connector

Female

Coaxial

Connector

Crystal

Holder Cavity

Connector

Advertising
This manual is related to the following products: