6 - 12

IP

N 07

4-

18

3X

XTC/C - XTC/2 Operating Manual

6.3.5 Leaf Spring Concerns

Spring conditions should be observed as part of the routine maintenance
interval. Insufficient bends or deformities in the spring contacts in the sensor
body are common causes of crystal problems. Lift each leaf spring up
approximately 60°. See

Figure 6-2

.

Figure 6-2 Shaping the Leaf Spring

60°

Leaf
Spring

Avoid kinking
leaf spring