INFICON XTC/C Thin Film Deposition Controller User Manual

Page 48

Advertising
background image

2 - 20

IP

N 07

4-

18

3X

XTC/C - XTC/2 Operating Manual

Table 2-2 Configuration Switch Settings

XTC/2

XTC/C

Switch 1

Test Mode

(0 = off, 1 = on)

Communications
Address

(2

4

)

Switch 2

Parameter Lock

(0 = off, 1 = on)

Communications
Address

(2

3

)

Switch 3

Control Mode

(0 = deposit, 1 = etch) Communications

Address

(2

2

)

Switch 4

Stop On Alarms

(0 = no, 1 = yes)

Communications
Address

(2

1

)

Switch 5

Stop on Max Power

(0 = no, 1 = yes)

Communications
Address

(2

0

)

XTC/C Switches 1-5 are only used for the optional IEEE488 (IPN 760-142-G1 or 757-122-G1).
[Addresses 0 to 30 are allowed.]

Switch 6

Recorder Type MSB

Communications Protocol
0 = INFICON, 1 = SECS

Switch 7

Recorder Type

Baud Rate MSB

Switch 8

Recorder Type LSB

Baud Rate LSB

NOTE: for the XTC/2
000 designates Rate, 100 Å/s full scale (unfiltered)
001 designates Rate, 1000 Å/s full scale (unfiltered
010 designates Thickness, 100 Å full scale
011 designates, 1000 Å full scale
100 designates Power %
101 designates Rate Deviation (±50 Å/s)
110 designates Rate 100 Å/s full scale - smoothed
111 designates Rate 1000 Å/s full scale - smoothed

NOTE: for the XTC/C
00 is 9600 baud
01 is 4800 baud
10 is 2400 baud
11 is 1200 baud

Switch 9

Beep On/Off

(0=on, 1=off)

Checksum

(0 = no, 1 = yes)

Switch 10

Backlight Dim

(0 = no, 1 = yes)

Unused

Switch 11

Start Layer without
backup crystal

(0 = no, 1 = yes)

Start Layer
without backup
crystal

0 = no, 1 = yes)

NOTE: See

section 4.16 on page 4-26

for description

Advertising