XTC/C - XTC/2 Operating Manual

3.7.2 Inputs

Inputs are activated by pulling the specific input's terminal to ground (<0.8V)
through a contact closure to common (GND) or with TTL/CMOS logic having
current sink capability of 2 ma (1 low power TTL load). These ports are read
every 250 ms; signals must be present during a read cycle.

Table 3-6 System I/O Connector

Input #

Pin #

Function

Description

13,14,15,

16,17

Input Common
(GND)

Used as reference for activating any of
the inputs

1

18

START deposition

Detection of a falling edge duplicates
front panel START

2

19

STOP deposition

Detection of a falling edge induces a
STOP

3

20

END deposit

Detection of a falling edge terminates
the Deposit state just as if the FINAL
THK were achieved.

Configuration switch #12 set for "Standard" Input Option:

4

21

SAMPLE INITIATE

Detection of a falling edge initiates a
RateWatcher sample if the film is
programmed for this feature.

5

22

SAMPLE INHIBIT

Application of a ground reference
voltage maintains the RateWatcher in
the Hold condition.

6

23

CRYSTAL FAIL
INHIBIT

Application of a ground reference
voltage prohibits the closure of the
Crystal Fail Relay and the associated
Stop.

7

24

ZERO thickness

Detection of a falling edge duplicates
the front panel ZERO.

8

25

SOAK 2 HOLD

Application of a ground reference
voltage extends the SOAK 2 state until
the signal/closure is removed.