INFICON HPG400-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual
Page 15
Advertising
tira03e1-b (0310) BPG/HPG400-SDv1.cp
15
This attribute contains the position of the hot cathode sensor calibration switch
(
→ 6).
This attribute contains the position of the Emission On threshold switch (
→ 6).
Depending on the value of the data type attribute, this attribute is INT or REAL
value.
Position of the
"Emission On"
Switch (P
↔HP)
Emission on
threshold
[mBar]
Emission on
threshold
[Torr]
Emission on
threshold
[Pa]
Emission on
threshold
[Counts]
0 or 1
1
0.75006168
100
28333.2625
2 or 3
0.5
3.75031×10
-1
50
28132.57634
4 or 5
0.2
1.50012×10
-1
20
27867.2837
6 or 7
0.1
7.50062×10
-2
10
27666.5975
8 or 9
0.05
3.75031×10
-2
5
27465.9113
Sensitivity
(Only HPG400-SD)
Emission On Threshold
(Only HPG400-SD)
Advertising
This manual is related to the following products: