INFICON HPG400-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual

Page 24

Advertising
background image

24

tira03e1-b (0310) BPG/HPG400-SDv1.cp

The setpoints can only be read by DeviceNet. It is not possible to set the
values or to influence the state of the relays by DeviceNet. Even if the de-
vice is not allocated, the relay status is set corresponding to the voltage ad-
justed by the two setpoint potentiometers.
Only in case of a microcontroller RESET (for example reset on Identity Ob-
ject), the relays will be reset too.

The setpoint is available in the actual pressure unit and data type. If the pressure
unit "Counts" is set, use the following formula for the conversion from Counts to
pressure:

BPG400-SD:

p

mbar

=

10

Counts / 2000 - k

where:

k

mbar

=

12.5

k

Torr

=

12.624903

k

Pa

=

10.5

The relation between setpoint and voltage is:

p

mbar

=

10

(U - 7.75) / 0.75 + c

where:

c

mbar

=

0

c

Torr

=

-0.125

c

Pa

=

2

HPG400-SD:

p

mbar

=

10

9 × Counts / 1024 - 6

The relation between setpoint and voltage is:

p

mbar

=

10

9 × U / 10 - 6

The setpoints are only activated, if the pressure is below 100 mbar, therefore a
setpoint above 100 mbar can not be realised.

The S-Analog Sensor Object provides the following Common Services:

Service

Code

Service Name

Description of Service

0E

h

Get_Attribute_Single

Returns the contents of the specified attribute.

10

h

Set_Attribute_Single

Modifies an attribute value.

Service

Code

Service Name

Description of Service

4C

h

Full Scale Adjust

Performs a Full Scale Adjust for the Pirani

There are no state transitions associated with the invocation of this service. It is,
therefore, incumbent upon the user to establish the device into the desired con-
figuration prior to, and during the execution of this service. This will generally in-
volve exposing the sensor to a known environment and treating the values read
during execution of the services accordingly.
A success service response indicates that the service was accepted and the ap-
plication process started.

2.4.3 Common Services

2.4.4 Object-Specific

Services on Instance 1 /
Pirani