SELF-TEST

EXCEPTION

SELF-TESTING

ABORT

CRITICAL FAULT

IDLE

EXECUTING

Self-Test Failed

Power Applied, or
Reset Request from any state except CRITICAL FAULT, or
Perform Diagnostics Request from any state except CRITICAL FAULT or ABORT

Self-Test

Passed

Exception Condition

Cleared

Critical Fault

from any state

Start

Request

Stop

Request

Recover Request

Abort Request

Abort Request

Abort Request

Recover Request or

Abort - Used to transition the device application objects to the aborted state. This
service request may be (and generally will) originated internally, from application
objects.
Recover - Used to transition the device application objects, out of the abort state,
to the idle state. This service request may be originated internally, from application
objects.
Perform_Diagnostics - Used to instruct the DS object to perform a diagnostic test.

Service
Code

Service Name

Description of Service

0E

h

Get_Attributes_Single

Returns the contents of the specified
attribute.

10

h

Set_Attributes_Single

Modifies an attribute value.

05

h

Reset

Resets the device to the Self-Testing state.

06

h

Start

Starts the device execution.

07

h

Stop

Moves the device to the Idle state.

The device transitions from the IDLE state to the EXECUTING state by a
START Request (Service Code 06

h

) or by the receipt of the first valid I/O

data!
You will not get any valid measurement values from the device until this
service has been requested or the I/O-poll message has been received.