INFICON HPG400-SD ATM to Ultra-High Vacuum Gauge (DeviceNet) User Manual

Page 22

Advertising
background image

22

tira03e1-b (0310) BPG/HPG400-SDv1.cp

Attribute

ID

Access

Rule

NV/
V

Name

DeviceNet
Data Type

Description of Attribute

Semantics of Values

96

Get

V

Status
Extension

BYTE

Bit mapped byte providing
additional status bits

Bit description:
0 Reading Invalid

(Logical Inversion of
Reading Valid)

1 Overrange Exceeded
2 Underrange Exceeded

99

Get

NV

Subclass

UINT

Identifies the subset of
additional attributes, serv-
ices and behaviors for hot
cathode ion gauges

5

Hot Cathode Ion Gauge

→ Instance 1

An S-Analog Sensor object instance derives a reading from a physical analog
sensor. The reading is converted to the data type and units specified for the Value
attribute.
Using Counts and INT the following conversion has to be used:

BPG400-SD:

Counts

=

[ log

10

(pressure) + k ] × 2000

where:

k

mbar

=

12.5

k

Torr

=

12.624903

k

Pa

=

10.5

HPG400-SD, Hot cathode measuring range 8333 < counts < 24333:

Counts

=

[ log

10

(pressure) + k ] × 2666.665

where:

k

mbar

=

9.125

k

Torr

=

9.249903

k

Pa

=

7.125

→ Instance 1

→ Instance 1

16 Bits are used as sensor faults . Bit 8

… Bit 16 are mapped to the Exception

Detail Alarm 3, Bit 0

… Bit 7 are mapped to the Exception Detail Alarm 2.

Data Component

Bit 7

Bit 6

Bit 5

Bit 4

Bit 3

Bit 2

Bit 1

Bit 0

Sensor Alarm Byte 0

0

0

0

0

0

0

0

0

Sensor Alarm Byte 1

0

0

0

0

0

0

Electronics

Failure

0

2.4.2.4 Semantics of S-Analog

Sensor Instance 2

Data Type

Value

Safe State

Safe Value

Sensor Alarm