1 Starting-Up of the Slave
The BPG400-SD / HPG400-SD has to be powered with two voltages:
1.)
24 Volt DC, 18 W at the 15 pole Sub-D connector for the gauge itself;
2.)
24 Volt DC nominal, <2 W range 11 … 25 V) at the DeviceNet micro style
connector for the DeviceNet transceiver.
1
2
3
4
5
6
7
Position
Function
1
Address Switch × 10, decimal
2
Address Switch × 1, decimal
3
DeviceNet connector
4
Data Rate Switch
5
Network Status LED
6
Module Status LED
7
"Sensor cable" connector
(Power, analog I/O, RS232C I/O
and Relay contacts)