JULABO FCW1600_1600S Recirculating coolers User Manual

Page 27

Advertising
background image

Recirculating coolers

27

6.4.

PID control parameters

For internal and external control two separate parameter
sets are available.
The PID control parameters can be adapted to the
requirements of the controlled member.
The values are preserved after switching off the
recirculating cooler.

SET



ACTUAL

SET





 The control parameters are indicated by operating the

key

(5.1).


Indications in case of internal control

INT:

Parameter Setting

range

CP 1

Xp (example 4.0 K).

0.1 ... 100 K

CP 2

Tn (example 160 s).

1 ... 9999 s

CP 3

Tv (example 20 s).

0 ... 500 s







Each indicated control parameter can be optimized
manually .

Setting

:

Operate the key

as often until the desired control

parameter is indicated. Example: CP3

Use the cursor keys

to move left or right on

the display until the numeral you wish to change is
blinking.

Use the increase/decrease arrows

to

change the selected numeral (-, 0, 1, 2, 3, ... 9).

Press enter

to store the selected value