Electrical connections – JULABO FCW1600_1600S Recirculating coolers User Manual

Page 30

Advertising
background image

Electrical connections

30

7.1.

Other error messages

ACTUAL

SET

Incorrect/invalid entry. Value too small or too large, or
function not available.

Under menu item E_Sb the parameter is set to 1, and
the connection between Pin 2 and Pin 3 of the stand-
by connector is interrupted (see page 30).


8. Electrical

connections

Notice:

Use shielded cables only.

The shield of the connecting cable is electrically connected to the plug housing.

Stand-by input (L)


1

2

3

AK

Pin

assignment:

Pin

Signal

1

not

used

2

5 V / DC

3

0 V


Activate the stand-by input:

Under menu item E_Sb, set the parameter to 1
(see page 36).
Connect an external contact 'AK' (e.g. for emergency switch-off)
or an alarm contact of the superordinated application system.
In case the connection between Pin 2 and Pin 3 is interrupted by
opening the contact 'AK', a complete shutdown of the circulating
pump, heater and cooling compressor is effected, and the unit
enters the condition "stand-by". The message "E_Sb" appears on
the MULTI-Display (LED2) (see page 30).

Alarm output (M)

1

2

3

This potential-free change-over contact is activated in case of an
alarm.
Pins 2 and 3 are connected under the following conditions:
-

alarm

- status "OFF" and "rOFF"

- mains switch "off"

Switching capacity max. 30 W / 40 VA
Switching voltage max. 125 V

/

Switching current max. 1 A