Remote control – JULABO FCW1600_1600S Recirculating coolers User Manual

Page 32

Advertising
background image

Remote control

32

9. Remote

control

9.1.

Communication with a PC or data system

ACTUAL

SET

For remote control, under the menu item OP (Operating
mode) set the parameter to 1.
The message "rOFF" appears on the display.
In general, the computer (master) sends commands to the
recirculating cooler (slave). The recirculating cooler sends
data (including error messages) only when the computer
asks for it.
A transfer sequence consists of:

 command

 space

(

; Hex: 20)

parameter (the character separating

decimals in a group is the period)

end of file (

; Hex: 0D)

The commands are divided into in or out commands.
in commands:

asking for parameters to be

displayed
out commands: setting

parameters


The out commands are valid only in remote control mode.

Examples:

 Command to set the setpoint to 5.5 °C:

out_sp_00

5.5

 Command to ask for the setpoint:

in_sp_00

 Response from the recirculating cooler:

Advertising