Recirculating coolers

33

9.2.

List of commands

Command

Parameter

Response of recirculating cooler

version

no

Number of software version

status

no

Status message (see below)

out_mode_04
out_mode_04
in_mode_04

0
1
no

Set control mode via PC
Set control mode via programmer input (O)
Ask for actual control mode

out_mode_05
out_mode_05
in_mode_05

0
1
no

Stop the recirculating cooler = rOFF
Start the recirculating cooler
Ask for actual condition (Start/Stop)

out_sp_00
in_sp_00

xx.x
no

Set working temperature value
Ask for working temperature value

in_sp_01

no

Ask for high temperature value

in_sp_02

no

Ask for low temperature value

out_sp_03

in_sp_03

xxx

no

Set control ratio for feed/return flow temperature
Ask for actual control ratio

in_pv_00

no

Ask for actual feed temperature

in_pv_01

no

Ask for actual temperature of external sensor

in_pv_02

no

Ask for actual heater capacity

in_pv_03

no

Ask for actual return temperature

in_pv_04

no

Ask for actual safety temperature

out_par_06
in_par_06

x.x
x.x

Control parameter Xp of the internal controller
out = set; in = ask

out_par_07
in_par_07

xx.xx
xx.xx

Control parameter Tn of the internal controller
out = set; in = ask

out_par_08
in_par_08

xx.xx
xx.xx

Control parameter Tv of the internal controller
out = set; in = ask