13

KDS Gemini 88

Valve Control Box

Setting up Valve

Set up consists of:

1)

Pump to Valve Electrical Connections:

The cable fitted with a male 9 pin D-sub connector must be connected to the

female D-sub connector on the rear of the Model 33 Syringe Pump. The Valve

Power can now be swiched on with the illuminated power On/Off switch on the

front valve housing.

2)

Valve tubing to syringe hook up:

The valve(s) are de-energized when the front mechanism (#1) is running in the

INFUSE direction. The syringe on the #1 mechanism when infusing should

therefore be connected to the lower tubing in the pinch valve (see figures on

next page).

Continuous Delivery Valve Boxes

Model No.

Description

55-7000

Pump 33 Continuous Delivery Valve Box, normal pressure, 30 p.s.i. supplied with

3.2 mm ID X 6.4 mm OD (1/8” ID X 1/4” OD) silastic tubing and a connector

cable for connection to the syringe pump.

55-7001

Pump 33 Continuous Delivery Valve Box, high pressure, 200 p.s.i. supplied with

6.4 mm (1/4) OD stainless steel tubing with SWAGELOK

®

fittings and connector

cable for connection to the syringe pump.