Appendix C: Command Information

PUMP CHAIN COMMAND SUMMARY

All spaces in command are ignored. Maximum 5 digits per integer, or float data.

command

= >

[<adr>] [<cmd>] <CR>

adr

= >

[d]d

1 or 2 digit address

cmd

=>

RUN

Start pump

STP

Stop pump

RAT [A] [<float> [<units> ] ]

Set/query syringe 1 rate

RAT B [<float> [<units> ] ]

Set/query syringe 2 rate

DIA [A] [<float>]

Set/query syringe diameter 1

DIA B [<float>]

Set/query syringe diameter 2

MOD [{AUT|PRO|CON}]

Set/query pumping mode

DIR [{INF|REF|REV}]

Set/query pumping direction

PAR [{ON|OFF}]

Set/query Parallel/Reciprocal

IN <d>

Read TTL pin d

OUT <d> = {ON|OFF}

Set TTL pin d

SAV

Save settings in NOVRAM

VER

Query software version

units

= >

{UM|UH|MM|MH}

µl/mn, µl/hr, ml/mn, ml/hr

COMMAND RESPONSE OUTPUT FORMAT:

output

= >

[<response>] <prompt>

response

= >

<response> <response>

<lf> <message> <cr>

prompt

= >

<lf> <adr> :

Stopped

<lf> <adr> >

Syringe 1 infusing

<lf> <adr> <

Syringe 1 withdrawing

<lf> <adr> *

Pump stalled

adr

= >

[d]d

1 or 2 digit pump address

message

= >

?

Unrecognized command

NA

Command not presently applicable

OOR

Entered value out of range

<text>

<integer>

<float>

All spaces ignored. Maximum 5 digits per integer or floating point number.

LEGEND

float

= >

f f f f f f

f

= >

d|.

integer

= >

d d d d d

d

= >

0|1|2|3|4|5|6|7|8|9

b

= >

space

cr

= >

(ASCII 13)

lf

= >

(ASCII 10)

text

= >

(any string of ASCII characters)

< . . . >

non-terminal

upper case

terminals

[ . . . ]

optional

{ . . . }

select one

|

either-or