Pump 33 requires no special maintenance other than keeping it clean by avoiding accidental

spills of pumped material.

The two guide rods and the lead screw should be sparingly lubricated periodically with the

Magnalube-G R grease provided with the pump. This

PTFE R based grease is available

either from KD Scientific or Carleton-Stuart Corp. 13-02 44th Ave., Long Island City, NY

11101.

Solvents of any type should never be used to clean the pump. A mild detergent solution may

be used to clean the keypad.

FUSE REPLACEMENT

Make sure power cord is disconnected from main supply before replacing fuse. Remove bot-

tom cover on pump by removing four (4) rubber feet and four (4) screws. Locate fuse on

power supply module. Remove fuse from fuse clip. Use caution no to break fuse when remov-

ing. Replace fuse, bottom cover, screws and rubber feet. For continued fire protection replace

fuse only with 250V fuse of the specified type and rating. (3AG 1AMP 250V SLO-BLO)

Figure 11. Fuse Replacement

WARNING:

Use in manner not specified by manufacturer may impair

protection provided by equipment.