Kds gemini 88 – KD Scientific Gemini 88 User Manual

Page 23

Advertising
background image

KDS Gemini 88

PUMP CHAIN COMMANDS AND RESPONSES

(continued)

DIR [{INF|REF|REV}]

Request to set or query pumping direction of syringe 1.

SET:

Command

Meaning

DIR INF

Sets syringe 1 pumping direction to infusion

DIR REF

Sets syringe 1 pumping direction to refill

DIR REV

Reverses pumping direction of syringe 1

Direction of syringe 2 is set appropriately.

Query pumping direction of syringe 1:

DIR

Response will be one of:

INFUSE

REFILL

PAR [{ON|OFF}]

Request to set or query Parallel or Reciprocal setting.

SET:

Command

Meaning

PAR ON

Sets syringes to parallel pumping direction.

PAR OFF

Sets syringes to reciprocal pumping direction.

Direction of syringe 2 changes appropriately, while direction of syringe 1 does not change.

Query:

PAR

Possible Responses

Meaning

ON

Parallel

OFF

Reciprocal

IN <d>

Request to read in the TTL logic level at the specified pin, “d”, on the external 9 pin

D-SUB connector.

Valid pin numbers for input are:

6, 7 and 8

Possible Responses

Meaning

ON

TTL level high

OFF

TTL level low

If the pin specified is invalid, a “Not Applicable” error is returned.