PUMP CHAIN COMMANDS AND RESPONSES

(continued)

DIA [{A|B}] [<float>]

Request to set or query syringe diameter settings.

Set syringe 1 diameter:

DIA diameter

DIA A diameter

Set syringe 2 diameter (only valid in proportional mode):

DIA B diameter

Corresponding syringe’s rate will be zeroed.

Diameter is of the format:

f f f f f f

Units are millimeters.

Diameter 2 is only accepted when mode is proportional. If diameter is accepted and valid,

diameter will become the new diameter. Diameter

will not

be accepted if the pump is run-

ning and a “Not Applicable” error is returned. If the diameter is invalid, an “Out Of Range”

error is returned.

Query syringe 1 diameter:

DIA

DIA A

Query syringe 2 diameter (only valid in proportional mode)

DIA B

Response is of the following format:

f f f f f f

Units are millimeters.

MOD [{AUT|PRO|CON}]

Request to set or query pumping mode

SET:

Command

Meaning

MOD AUT

Puts pump in Auto Stop mode

MOD PRO

Puts pump in Proportional mode

MOD CON

Puts pump in Continuous mode

If pump is running, command will not be accepted and a “Not Applicable” error is returned.

Query:

MOD

Possible Responses

Meaning

AUT

Auto Stop mode

PRO

Proportional mode

CON

Continuous mode