Kds gemini 88, Specifications – KD Scientific Gemini 88 User Manual

Page 4

Advertising
background image

2

KDS Gemini 88

Specifications

Type

Microprocessor dual drive, single syringe, infuse/withdraw

Accuracy

±0.35%

Reproducibility

±0.1%

Syringe:

Type

Plastic, glass or stainless steel

Size Minimum

0.5 µl

Size Maximum

140 ml

Flow Rate:

Minimum

0.0004 µl/hr

Maximum

106.6 ml/min

Non Volatile Memory

Storage of all settings

RS-232

RJ11-4 conductor

TTL

9-pin connector

Average Linear Force

57 lbs

Drive Motor

2 motors, each 0.9° step angle motor

Motor Drive Control

Microprocessor controlled from 1/2 to 1/4

microstepping

Motor Steps per

1,600 steps at 1/2 stepping or 3,200 steps at

Rev. of Leadscrew

1/4 stepping

Step Resolution

0.33 µm/step

Step Rate:

Minimum

27.3 sec/step

Maximum

416.7 µsec/step

Pusher Travel Rate:

Minimum

0.726699 µm/min

Maximum

95.25 mm/min

Power

45 W, 1.0 A fuse

Voltage Range

95 to 130 VAC, 60 Hz; 220 to 260 VAC, 50 Hz,

selectable

Dimensions, H x W x D

15.2 x 31.1 x 28.6 cm (6 x 12.5 x 11.25 in)

Weight

6.8 kg (15 lb)

Timing Belt Drive

2:1

Lead Screw Pitch

24 threads/inch

Display

5 digits plus 14 indicator LEDS