Instructions for Use V1.3 – 10/2012

2.

Optional components

Premium blade holder (for low‑profile and

high‑profile blades) with anti‑roll system

Assembling the anti‑roll system for the Pre‑

mium blade holder
1. Insert the glare-minimizing glass insert into the

interchangeable frame and tighten it evenly

using the knurled screw (10).

2. Insert the shaft (11) of the metal frame for ex-

changeable glass inserts from above into the

bore of the swinging arm (12) in such a way

that the pin rests in the notch.

3. Push the white plastic plate (14) from below

onto the shaft (11).

4. Screw the knurled nut (13) from below onto

the shaft (11).

13

14

11

10

10

Fig. 13

• Pressure plate with adjustable anti-roll system

and glass inserts for various section thick-

nesses

• Blade ejector (1) and knife guard (2)
• Integrated lateral movement and stable base
• Clearance angle adjustment (5) with 4 mm

Allen key (see detail image, to left at blade

holder) - recommended angle from 2° to 5°.

• Lever (6) for the lateral movement
• Lever (7) for clamping the blade
• Insert for low-profile blades (7a, Fig. 14) for

insertion into Premium blade holder.

1

5

2

3

4

6

7

8

12

Fig. 12

All four sides of the glass anti‑roll plate

can be used. Replacement glass can be

reordered when necessary.

Red elements on the blade holders,

such as the knife guard and ejector,

are protective devices that must not be

removed.