Clearance angle adjustment – Leica Biosystems Premium blade holder User Manual

Page 9

Advertising
background image

9

Leica Premium blade holder

1. Sectioning – Premium blade holder

Clearance angle adjustment

1. The clearance angle scale is located on the

left side of the blade holder.

2. Release the blade holder by turning the No.

4 Allen screw (23) counterclockwise. Select

a clearance angle of 0°. To do so, align the

number 0 to the index mark (25) and tighten

the Allen screw (23). If the sectioning results

are not satisfactory, increase the clearance

angle in 1° increments until you achieve op-

timum results.

25

23

• A clearance angle that is too steep or too shallow may lead to poor sectioning results

and can possibly damage the specimen.

Fig. 11

The position of the knife edge relative to the specimen will change when the clearance angle is

adjusted. For this reason, always place the specimen above the knife when adjusting the clear‑

ance angle. Otherwise, the specimen could hit the knife while being raised. Carefully check the

distance between the block face and the cutting surface before taking the first section.

Settings of 2° ‑ 5° (premium blade

holder) are well suited for most appli‑

cations.