Removing the blades, Cutting – premium blade holder – Leica Biosystems Premium blade holder User Manual

Page 7

Advertising
background image

7

Leica Premium blade holder

12

Magnet

10

4

Inserting low‑profile blades (continued)

Two magnets are attached to the rear side of the

insert. These point away from the operator after

the insert has been inserted (towards the rear

pressure plate).

Then insert the blade as described (for high-

profile blade).

Removing the blades

1. Ensure that the handwheel is securely locked.
2. Fold the anti-roll system (4) to the left – while

doing so, hold the lever (11) (not the adjusting

screw of the anti-roll system) so that the height

of the anti-roll system remains unchanged.

3. Open the clamping lever (10) by turning it

counterclockwise (Fig. 7).

4. Push the blade ejector lever (13) so that it

contacts the left side of the blade and pushes

it to the right.

5. Use the magnet on the end of the Leica brush

(12) to attach to the exposed right side of the

blade.

6. Carefully lift out the blade (9).

1. Cutting – premium blade holder

Fig. 6

Fig. 7

Fig. 8

10

9

4

13

11