Daily cleaning of the premium blade holder – Leica Biosystems Premium blade holder User Manual

Page 13

Advertising
background image

13

Leica Premium blade holder

5. Afterwards, the pressure plate (6) can be removed for cleaning (with alcohol).
6. To remove the upper part (8) from the segment arc (2), release the blade holder clamping lever (6,

Fig. 12) by rotating it clockwise and pulling it out. The blade holder can now be moved sideways

until it slides off the segment arc.

7. To disinfect, bring the parts to room temperature and either soak them in disinfectant or place on

an absorbent cloth and spray them liberally and uniformly with disinfectant. After the recommended

exposure time, dry completely before placing in the cryochamber.

3.

Daily cleaning of the premium blade holder

Fig. 16

6

3

4

5

7

8

1

1. To release the segment arc (1) from the

base, use a 4 mm Allen key to loosen the

screw over the clearance angle adjust-

ment (2) and remove the segment arc (1)

from the blade holder base.

2. Fold the anti-roll system (3) to the left using

the lever (4).

3. Open the clamping lever (5) by turning it

counterclockwise.

4. Hold pressure plate (6), and remove bolt

(7).

For daily cleaning, remove the section waste from the blade holder using a dry brush. Please use

a cold brush, as otherwise the section waste will thaw and stick to the blade holder.

Mechanical damage to the pressure plate will strongly affect the sectioning quality. For this

reason it is important to ensure that the clamping area is not damaged during cleaning or other

work.

For disinfection, commercially available, mild cleaners and disinfectants can be used – we rec‑

ommend Leica Cryofect (not available in the USA).

Wear gloves during cleaning to prevent frostbite.

If several blade holders are cleaned at the same time, the parts must NOT be mixed up! Failure

to adhere to this may result in sectioning problems!

2