Lateral movement – Leica Biosystems Premium blade holder User Manual

Page 8

Advertising
background image

8

Instructions for Use V1.3 – 10/2012

Fig. 9

7. Once the blade has been removed from the

blade holder, dispose of it in the dispenser

container (storage compartment in bottom,

Fig. 9) or by the method compliant with the

institutional guidelines.

1.

Sectioning – Premium blade holder

Lateral movement

If the sectioning results are not satisfactory, the

blade holder (on the segment arc in figure 10 )

can be shifted sideways in order to use another

part of the blade, and to benefit from the entire

length of the blade.

1. Turn the clamping lever (14) clockwise to re-

lease, and then shift the blade holder sideways

to the desired position.

2. To tighten, turn the clamping lever (14) coun-

terclockwise.

Fig. 10

14

Guide the brush with the magnet to the

blade and lift it upwards and out. The

cut‑resistant safety gloves included in

standard delivery must be worn when

disposing of the blade!