Inserting blades into premium blade holder, Inserting high-profile blades, Inserting low-profile blades – Leica Biosystems Premium blade holder User Manual

Page 6

Advertising
background image

6

Instructions for Use V1.3 – 10/2012

Fig. 3

Fig. 4

1.

Sectioning – Premium blade holder

Inserting blades into premium blade holder

10

9

4

11

Inserting high‑profile blades

1. Fold the anti-roll system (4) to the left – while

doing so, hold the lever (11) (not the adjusting

screw of the anti-roll system) so that the height

of the anti-roll system remains unchanged

(Fig. 3).

2. Open the clamping lever (10) by turning it

counterclockwise (Fig. 3).

3. Carefully insert the blade (9) from above or

from the side between the pressure plate and

the insert for low-profile blades. Make sure

that the blade is inserted so that it is centered

and lies evenly on the edge (see red arrow in

Fig. 3).

4. Clamp the clamping lever (10) by turning it

clockwise (see Fig. 4).

5. Fold the anti-roll system (4) back to the right

(toward the blade) using the lever (11).

Inserting low‑profile blades

• When using low-profile blades, the insert for

low-profile blades must first be placed in the

blade holder. Then insert the blade.

4

11

10

Fig. 5

Caution!

Microtome blades are extremely sharp!

The premium blade holder can be used for both low‑profile and high‑profile blades.

The safety gloves included with stan‑

dard delivery must be worn when in‑

serting the blade!

The anti‑roll system functions as a

knife guard here!