5.1.2 Changing the wax

Draining the wax

Connect the paraffin drain hose (2) as shown
in

Fig. 13

and insert the loose end of the hose

into a remote collecting container.

Fig. 13

Select the wax bath to be drained and the
remote container (‘Remote’).
Then press STATION DRAIN.

Several queries will be displayed each of
which has to be confirmed individually.

1

2

From the FAVORITES screen, press SMART
SCREEN (see previous chapter).

Make sure the hose end is securely kept in
the remote container until both the drain has
been completed and the hose has been
purged with pressurized air.

+