Operation – Leica Biosystems ASP300 User Manual

Page 33

Advertising
background image

33

Leica ASP300

Fig. 16

→

→

Adding molten paraffin

•

Molten wax may have a maximum temperature of 70 ° C.

•

If filling with molten wax, do not fill above the MAX level (see

Fig. 15

).

•

After a wax bath has been filled, it must be marked as ‘full’ in the
REAGENT STATUS screen.

Software control

•

If the wax baths are filled with molten wax, it is possible to override
the Leica ASP300 melting time calculation.

For that purpose, in supervisor mode, access the
SERVICE FUNCTIONS screen (

Fig. 16

).

•

Press TIME UNTIL WAX MELTED.

•

Enter the required delay time and press OK
to confirm.

If the instrument attempts a paraffin step with the paraffin only partly molten, severe instru-
ment malfunctions may occur.

→

→

5.

Operation