4 preventative maintenance schedule, Preventative maintenance schedule, Cleaning and maintenance – Leica Biosystems ASP300 User Manual

Page 67

Advertising
background image

67

Leica ASP300

7.

Cleaning and maintenance

Maintenance activity

daily

weekly

monthly

every

4 months

Lubricate reagent bottle O-rings

√

√ ∗

and check for damage.

Wipe dry retort and lid

√

after completing retort clean cycle.

Clean exterior of instrument

√

with soft cloth and minimal
amount of xylene.

Check retort strainer

√

for any tissue or wax debris.

Remove residual paraffin

√

from inner surface of wax bath lid.

Inspect and, if necessary,

√

clean retort lid seal.

Inspect and, if necessary,

√

clean wax bath lid seal.

Inspect and, if necessary,

√

adjust paraffin levels.

Inspect reagent container fluid levels.

√

7.4 Preventative maintenance schedule

*

if the automatic fill and drain function is actually being used