38

Instruction manual V 1.3 – 12/2003

Allocation of reagents to the correct reagent group is the basis for
compatibility monitoring. Allocation to the wrong reagent group
can lead to reagent cross contamination.

Entering / modifying reagent thresholds

•

Highlight the reagent to be modified.

•

Press the headline key symbol of the parameter to be changed.

•

Enter a new threshold or - if no warning is desired - press CLEAR to
remove the threshold altogether.

•

Press OK to confirm.

The same threshold value applies for all reagent stations con-
taining the same reagent.

Changing reagent names or reagent groups

•

Highlight the reagent.

•

Press the corresponding headline key symbol.

•

In the entry window (or via the keyboard), enter the new reagent
group allocation.

•

Press OK to confirm and save.

•

Reedit all station and programs allocated to the reagent changed.

5.

Operation

Changeable parameters