24

Adjusting Environmental Chamber Position

Installation

Adjusting Environmental Chamber Position

1. It is necessary to adjust the position of environmental chamber via upper

and lower pull rods on site.
Before fixing bracket and environmental chamber, slide the chamber from
front to back on the bracket, and adjust the position of the bracket from left
to right on testing machine to make the pull rods align to the center of the
rod hole of environmental chamber.

2. After the pull rods align to the center of pull rod hole of environmental

chamber, fix the bracket to the testing machine, and fix the stoppers to the
front bracket.

Rod Hole

Adjust position
of bracket.

Oriented Stops

Align rod to center
of rod hole.

Adjust position
of cabinet.

Fix the stoppers to locate the
front stopping point of
environmental chamber.