Connecting Environmental Chamber Mains Power

Installation

Connecting Environmental Chamber Mains Power

Power connection

For MTS series FEC chamber use the 3-wires cable that is provide for power
input, and connect to the customer electrical box and disconnect.

Note

Local electrical codes supersede any information found here:

Voltage Single-phase AC 200-240
Current 15Amps
Frequency 50 Hz/60 Hz

Note

Electrical connections must be made by qualified personnel and is their
responsibility for using the proper power disconnect that conforms to
local codes and regulations when connecting the machine to the building
main power.

Electrical disconnect

The customer is responsible for providing an electrical power disconnect that is
easy to operate and easy to reach. It must also meet IEC 60947-1 and IEC 60947-
3 standards. Recommended circuit breaks would be ones that are of the thermal
magnetic type with characteristics suitable for large inductive loads (D-type trip
characteristic). If fuses are used it is recommended that they are of the time delay
type with dual elements. These recommendations should be followed to avoid
nuisance tripping.