Adjusting Environmental Chamber Position

Series FEC Environmental Chamber Product Manual

Installation

25

3. In order to prevent the danger of Chamber overturn, FEC serials chamber

use four anti-overturn blocks to connect the chamber body and the rail
bracket. Please fix the four anti-overturn blocks on the left and right side
wheel of the chamber after the chamber installation and adjustments are
done.

4. To ensure the safety of the chamber operation, a door switch is installed on

the bottom of it. When the door is open, the heating or cooling function will
be cut off and the stirring fan will stop working at the same time. After you
closed the door, the stirring fan will automatically run again, but the heating
or cooling function will not automatically go back to normal use until you
press "Enable" button again. After the chamber installation is done, please
make sure the door switch function works properly.

Anti-overturn Block

Door Switch