Series FEC Environmental Chamber Product Manual

Routine Maintenance

44

Do not open the chamber immediately after a test to prevent damage or
personal injury caused by high or low temperature.

To extend the service life of an environmental chamber, eliminate water
vapor in the cabinet or other residue left in the chamber after test.

Do not use corrosive organic solvents when cleaning chamber.

During test or after test, do not use any kind of shelter to cover the chamber
so as to avoid the case that the heat or water vapor can not be diffused and
damage the chamber.

Troubleshooting

Troubleshooting

S

YMPTOMS

R

EASON

S

OLUTION

Set Value window flashes and
displays “S.br” .

Measure signal input
exceeds range.

1. Inspect if specification setting of
sensor is correct.

2. Inspect if signal input is open circuit.

Temperature increase speed is
slower than original speed

Parameter setting may be
incorrect.

Inspect if setting of temperature control
parameter is correct.

Temperature rises out of
control

parameter setting is
incorrect.

Inspect if setting of temperature control
parameter is correct.

Temperature fails to mains at
setting temperature

Control parameter is
imperfect or lost.

Start self-adjustment to find out proper
control parameter.