Routine maintenance, Routine maintenance overview checklist, Routine maintenance 43 – MTS Series FEC Environmental Chamber User Manual
Page 43: Routine maintenance overview checklist 43, Recommended service
Routine Maintenance
Series FEC Environmental Chamber Product Manual
Routine Maintenance
43
Routine Maintenance
Routine Maintenance Overview Checklist
Recommended Service
C
ALENDAR
T
IME
U
SING
8 H
OURS
R
UNNING
T
IME
R
ATE
PER
D
AY
D
AILY
W
EEKLY
A
NNUALLY
R
UNNING
T
IME
-H
OURS
8 40
2000
Clean the chamber inside/chamber surface area
X
*
* Denotes services performed by equipment operators. Most of these procedures involve visual checks
that should not interfere with testing system operation. These checks are also completed by trained
field service engineers on each Routine Maintenance visit.
Check the door limit and E-stop
X
Check the cable connection
X
Check all the buttons function
X
System Inspection
Check the cable connection
MTS
System Checks
Check E-stop
MTS
Check door switch
MTS
Check rail bracket level
MTS
Check over-temperature alarm instrument
function
MTS
Lubrication
Stirring fan bearing
MTS