Frame location and ventilation, Environmental requirements – MTS Acumen Electrodynamic Test System Load Frame User Guide User Manual

Page 34

Advertising
background image

Frame Location and Ventilation

To ensure proper ventilation, locate the load frame approximately 300 mm (12 inches) from adjacent walls
and equipment. Allow approximately 1 m (3 feet) behind the equipment for service access. Do not block the
vent holes on the sides or back of the frame.

For comfortable working conditions and proper equipment operation, heat dissipation of the equipment must
be considered in providing adequate heating or air conditioning in the laboratory area. Heat dissipation can
be approximated by summing the heat losses going into a room (1 kVA is equivalent to 860 kcal/hr [3,400
Btu/hr]) and the gains from other sources such as furnaces and personnel.

Environmental Requirements

All MTS Acumen load frames are intended for indoor use only. This indoor environment must conform to the
following environmental specifications.

Environmental Requirements

Specification

Parameter

5° to 40° C (41° to 104° F)

Operating Temperature

13

5 to 85% Non-condensing

Operating Humidity

-18° to 49° C (0° to 120° F)

Storage Temperature

90% Non-condensing

Maximum Storage Humidity

2000 m (6562 ft)

Maximum Altitude

2

Pollution Degree

IP20

IP Rating

II

Overvoltage Category

13

Take care when using heat producing accessories, such as ovens. Emitted heat from such devices may
damage the MTS Acumen if not properly controlled.

34 MTS Acumen™ Electrodynamic Test System

Installation

Advertising